Calibration Procedures Calibration<br />

5.5.5 Paw Sensor<br />

1. 0 cmH 2 O Calibration<br />

FIGURE 5-6 Removing the 3L bag<br />

a. Remove the 3L bag from the breathing circuit.<br />

b. Set the APL Valve to 0 cmH 2O.<br />

c. Select Continue to start calibration.<br />

d. Follow the on-screen instructions.<br />

5 - 10 0070-10-0683 AS3000 Service Manual