28.02.2013 Views

Datascope - Mindray

Datascope - Mindray

Datascope - Mindray

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Installation Guide Installation Checkout Procedure<br />

15. Oxygen Sensor Calibration<br />

NOTE: Oxygen Sensor Calibration can be performed in all<br />

ventilation modes.<br />

NOTE: See ‘‘Periodic Maintenance Schedule of Service Activities’’<br />

on page 6-2 for when to calibrate the oxygen sensor.<br />

a. Allow the breathing system to warm up and reach thermal equilibrium<br />

(approximately 30-60 minutes).<br />

b. Press the MENU key and then use the Navigator Knob to scroll to the<br />

Calibrate menu tab (see FIGURE 2-13). Select the Start Calibration button.<br />

FIGURE 2-13 Calibrate Menu Tab<br />

c. After the Start Calibration button has been selected, the screen shown in<br />

FIGURE 2-14 or FIGURE 2-16will be displayed, instructing the user to remove the<br />

oxygen sensor from the Breathing System and expose it to room air for at least 3<br />

minutes before proceeding. The O 2 sensor voltage is displayed during the<br />

calibration. This is the amplified O 2 cell voltage at the A/D converter for the oxygen<br />

sensor.<br />

NOTE: The O 2 sensor voltage is not displayed for UI versions 2.24<br />

and lower.<br />

d. Flush the O 2 sensor with air from the auxiliary output for 5-10 seconds to ensure that<br />

no O 2 bubbles are trapped in the sensor.<br />

NOTE: Do not shake the O 2 sensor during calibration.<br />

NOTE: Keep the O 2 sensor in a vertical position, connector side up,<br />

during calibration.<br />

NOTE: Place the O 2 sensor on top of the heated block during<br />

calibration to minimize the temperature difference from<br />

within the heated block.<br />

AS3000 Service Manual 0070-10-0683 2 - 17

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!