Installation Guide Installation Checkout Procedure<br />
23. Verify that the drive gas pressure loss alarm signals activate and N2O stops and AIR continues to flow.<br />
a. Set the O2 flow for 1 L/min.<br />
b. Set the N2O flow to 1 L/min.<br />
c. Set the AIR flow to 1 L/min.<br />
d. Disconnect the O2 line supply to the AS3000, and Close the O2 cylinder.<br />
e. Verify that the flow of N2O stops when O2 is lost.<br />
f. Verify that AIR continues to flow.<br />
g. Verify the “O2 Supply Failure” message appears on screen, a steady audio tone<br />
should sound.<br />
h. Reconnect O2 line pressure source to the AS3000.<br />
i. Verify that the alarms are reset.<br />
j. Verify N2O flow has resumed once O2 pressure returns.<br />
24. Verify that the battery operation and charging icons appears.<br />
a. Set the vent mode to Standby.<br />
b. Disconnect AC line cord.<br />
c. Verify that the battery icon appears on the screen and indicates the charge level of<br />
the battery.<br />
d. Verify that the line cord icon is flashing.<br />
e. Verify that the flow meter backlight is illuminated during battery operation.<br />
f. When the battery icon is in a not full state, reconnect the AC line cord.<br />
g. Verify that the battery charge icon appears on the screen.<br />
h. Verify that the flow meter backlight is illuminated during AC voltage operation.<br />
25. Verify operation of the work light.<br />
a. Turn on the work light located on the bottom side of the top panel.<br />
b. Verify that it lights in both on positions.<br />
26. Verify that the Auxiliary O 2 and AIR Flowmeters operates.<br />
a. Verify an AIR flow of 15 L/min can be obtained by connecting the auxiliary AIR hose<br />
to the pressure source and opening the flow meter.<br />
b. Verify an O 2 flow of 15 L/min can be obtained by connecting the auxiliary O 2 hose<br />
to the pressure source and opening the flow meter.<br />
27. Verify that the Suction Regulator operates in LINE and REGULATE<br />
positions.<br />
a. Set the suction regulator's selection dial to LINE.<br />
b. Verify maximum suction vacuum on the regulator's gauge.<br />
c. Set the suction regulator's selection dial to REGULATE.<br />
d. Verify the suction vacuum is adjustable on the regulator's gauge.<br />
e. Set the suction regulator's selection dial to OFF.<br />
AS3000 Service Manual 0070-10-0683 2 - 23