28.02.2013 Views

Datascope - Mindray

Datascope - Mindray

Datascope - Mindray

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Preoperative Checklist Periodic Maintenance<br />

19. Test the Tidal Volume in the PCV ventilation mode:<br />

a. Set the O 2 flow to 2 L/min and set the N 2 O and AIR flow rates to minimum flow.<br />

b. Set the ventilator controls to:<br />

ATTRIBUTE SETTING<br />

Patient Type Adult<br />

Ventilation Mode PCV<br />

Target Pressure - PTARGET 20<br />

Breathng Rate - freq 8<br />

I:E Ratio - I:E 1:2<br />

PEEP - PEEP Off<br />

Inspiratory Slope - Tslope 0.5<br />

c. Select PCV to begin ventilation.<br />

d. Verify that the PEAK Pressure settles within ±4 of the set value within 3 breaths.<br />

e. Re-activate CMV ventilation mode.<br />

20. Low FiO 2 Alarm Test<br />

a. Set the low FiO 2 Alarm limit to 50% O 2 .<br />

b. Set the AIR flow control valve to 5 L/min.<br />

c. Set the FiO 2 flow controller to minimum flow.<br />

d. Verify the following Low FiO 2 alarm signals activate, within three ventilation cycles:<br />

Low FiO 2 message appears on the screen.<br />

An alarm tone sounds.<br />

e. Set the Low FiO 2 alarm limit to 18%.<br />

Verify the alarm signals stop activating.<br />

21. High FiO 2 Alarm Test<br />

a. Set the high FiO 2 Alarm limit to 49% O 2 .<br />

b. Set the FiO 2 flow control valve to 5 L/min.<br />

c. Set the AIR flow controller to minimum.<br />

d. Verify the following High FiO 2 alarm signals activate:<br />

High FiO 2 message appears on the screen.<br />

An alarm tone sounds.<br />

e. Set the high FiO 2 alarm limit to the max setting.<br />

Verify the alarm signals stop activating.<br />

6 - 44 0070-10-0683 AS3000 Service Manual

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!