Calibration Calibration Procedures<br />

NOTE: If the system is going to be used during the calibration,<br />

insert the O 2 cell plug into the port from which the oxygen<br />

sensor was removed using a push and turn motion.<br />

a. Select 21% to enter the 21% oxygen concentration calibration screen.<br />

b. Remove the oxygen sensor from the Breathing System and expose it to room air for<br />

at least 3 minutes.<br />

c. Flush the O 2 sensor with air from the auxiliary output for 5-10 seconds to ensure that<br />

no O 2 bubbles are trapped in the sensor.<br />

d. Select Next to start 21% oxygen concentration calibration.<br />

NOTE: The O 2 sensor voltage is displayed during the calibration.<br />

This is the amplified O 2 cell voltage at the A/D converter for<br />

the oxygen sensor. The O 2 sensor voltage is not displayed<br />

for UI versions 2.24 and lower.<br />

e. When calibration is successfully completed, install the oxygen sensor into the<br />

Breathing System.<br />

NOTE: The oxygen sensor must be installed in the Breathing<br />

System for 10 minutes prior to 100% oxygen concentration<br />

calibration to adjust to the temperature of the system.<br />

f. Press the O 2 Flush button for 5-10 seconds to clear out any non-O 2 gases from the<br />

system.<br />

g. Expose the oxygen sensor to 100% pure oxygen (5 L/min) for at least 3 minutes.<br />

h. Select 100% to enter the 100% oxygen concentration calibration screen.<br />

i. Select Next to start 100% oxygen concentration calibration.<br />

AS3000 Service Manual 0070-10-0683 5 - 5