Installation Checkout Procedure Installation Guide<br />
c. Select Continue to prompt to the Normal Screen.<br />
FIGURE 2-12 Compliance Test Results<br />
14. Manual Leak Test<br />
NOTE: The Manual Leak Test detects smaller leaks than can be<br />
detected in the Automatic Leak Test.<br />
a. Ensure that the gas pressure for O 2 , N 2 O, and AIR are at 50 ±10 psi.<br />
b. Power ON the AS3000.<br />
c. Attach a reusable-silicone-rubber breathing circuit to the Breathing System.<br />
NOTE: For testing purposes always use a reusable breathing<br />
circuit.<br />
d. Tightly connect the Y-fitting on the breathing circuit to the test port.<br />
e. Attach a breathing bag to the bag arm.<br />
f. Set the APL Valve to the fully closed position (70 cmH 2 O).<br />
g. Rotate the O 2 Flow Control Valve until 50 cmH 2O pressure is observed on the<br />
Airway Pressure Gauge.<br />
h. Verify that the flow required to stabilize the pressure is less than 300 mL/min.<br />
2 - 16 0070-10-0683 AS3000 Service Manual