Datascope - Mindray
Datascope - Mindray
Datascope - Mindray
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Theory of Operation Ventilator Control and Drive<br />
Inspire Valve, J12<br />
PIN NAME FUNCTION<br />
1 - 2 24V 24V<br />
3 - 4 Valexp inspire Valve Drive<br />
Man/Auto Valve, J14<br />
PIN NAME FUNCTION<br />
1. 24V 24V<br />
2. Valman Manual/Auto Valve Drive<br />
Signals and Power of Sensor Board, J13<br />
PIN NAME FUNCTION<br />
1. A+12V Analogue +12V<br />
2. FL_INS Import flow<br />
3. FL_EXP Export flow<br />
4. FM_FL Flowmeter flow<br />
5. Pair Pressure of air<br />
6. Paw Pressure inside air way<br />
7. Pgas Pressure gas supply status<br />
8. GND Ground<br />
9. AGND Analog Ground<br />
10. AGND Analog Ground<br />
11. P24V Power 24V<br />
12. P24V Power 24V<br />
13. P24V Power 24V<br />
14. P12V Power 12V<br />
15. P12V Power 12<br />
16. P5V Power 5V<br />
17. P5V Power 5V<br />
18. Inhale Inspire Valve<br />
19. Man MANUAL/AUTO Valve<br />
20. Exhale Expire Valve<br />
21. Exhale Expire Valve<br />
22. AGND Analog Ground<br />
23. AGND Analog Ground<br />
24. AGND Analog Ground<br />
25. AGND Analog Ground<br />
AS3000 Service Manual 0070-10-0683 1 - 25