28.02.2013 Views

Datascope - Mindray

Datascope - Mindray

Datascope - Mindray

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

AS3000 Service Manual 0070-10-0683 3 - 9<br />

MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION<br />

“Ventilator setting is not<br />

possible!”<br />

“Displayed parameter values<br />

are for the pending mode”<br />

“Pressure, Volume and<br />

Apnea Alarms are OFF!”<br />

“Automatic Ventilation not<br />

available!”<br />

“BDU Communication<br />

Failure!”<br />

Runtime 1 A parameter has been set to a limit that can<br />

be set.<br />

Runtime 1 The AS3000 is in the pre-selection mode,<br />

but it is still working according to the<br />

original operating mode.<br />

Runtime 1 The functional alarm in Manual mode is<br />

disabled.<br />

Startup Test<br />

Runtime<br />

Startup Test<br />

Runtime<br />

1 A failure has been found (Displayed on the<br />

screen); the AS3000 can work only in<br />

Standby/Manual mode. This information is<br />

displayed after entering the Main Screen.<br />

1 The communication cable is disconnected.<br />

2 The BDU software has crashed or reset.<br />

“O 2 Sensor failure” Runtime 1 The main unit-oxygen sensor cable is<br />

disconnected.<br />

1 The parameter set cannot exceed the<br />

adjustable range. If the limit is not reached,<br />

another parameter may be limiting the range<br />

and needs to be adjusted accordingly<br />

1 Set parameters 15 seconds in advance; or<br />

directly confirm and enter ventilation mode; or<br />

press the NORMAL SCREEN button to return<br />

to the original ventilation mode interface; or do<br />

not perform any operation and it will<br />

automatically return to the original ventilation<br />

mode interface after 15 seconds.<br />

1 Enable the functional alarm in Manual mode.<br />

1 Handle it according to specific information on<br />

the failure.<br />

1 Check the communication cable.<br />

2 Power on again.<br />

1 Check the unit-oxygen sensor cable.<br />

“O 2 Supply failure” Runtime 1 The gas source is closed or disconnected. 1 Check the gas source.<br />

Continuous Pressure Runtime 1 Airway pressure is too high.<br />

2 The expiration valve (PEEP) is defective.<br />

3 The airway is occluded.<br />

1 Relieve the airway pressure<br />

2 Repair/replace the expiration valve<br />

3 Clear the occlusion<br />

High Airway Pressure Runtime 1 Airway pressure is high. 1 Adjust the upper limit set for the airway<br />

pressure alarm.<br />

Low Airway Pressure Runtime 1 Airway pressure is low. 1 Adjust the lower limit set for the airway<br />

pressure alarm.<br />

Negative Pressure Runtime 1 Airway pressure is continuously negative. 1 Check the airway and fresh gas setting<br />

Low FiO 2 Runtime 1 Oxygen concentration is lower than the<br />

lower limit set for Oxygen Concentration<br />

alarm.<br />

2 Oxygen calibration is inaccurate.<br />

1 Adjust the lower limit set for the Oxygen<br />

Concentration alarm or the actual oxygen<br />

concentration.<br />

2 Calibrate the oxygen concentration sensor<br />

again.<br />

Repair Information Troubleshooting Chart

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!