28.02.2013 Views

Datascope - Mindray

Datascope - Mindray

Datascope - Mindray

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Functional Tests Periodic Maintenance<br />

6.8.10.2 Vaporizer Accuracy Test<br />

WARNING: Avoid exposure to respiratory gases by always directing<br />

the fresh gas flow from the fresh gas outlet to the waste gas<br />

scavenger.<br />

1. Remove the Fresh Gas Hose from the Common Gas Outlet.<br />

2. Insert an endo-tracheal tube mask elbow adapter with monitoring port, into the Common<br />

Gas Outlet.<br />

3. Set the APL Valve to 70 cmH 2 O.<br />

4. Remove the Waste Gas Scavenger Hose from the bottom of the Breathing System EVAC<br />

port. Leave the connection to the waste gas scavenger attached.<br />

5. Set the waste gas scavenger flow above the minimum setting.<br />

6. Adapt the Waste Gas Scavenger Hose to connect to the endo-tracheal tube mask elbow,<br />

now connected to the Common Gas Outlet port.<br />

7. Attach the Agent Analyzer meter sampling tube to the endo-tracheal tube mask elbow's<br />

monitoring port.<br />

8. Fill the vaporizer with anesthetic agent (if necessary).<br />

NOTE: Do not overfill by filling past the indicator line on the<br />

vaporizer.<br />

9. Test the vaporizer accuracy per the Dräger Vapor 2000 instructions (See<br />

section 6.8.10.4), or see the appropriate vaporizer manual for testing details.<br />

10. Test each vaporizer in turn.<br />

11. Test any vaporizer on the Vaporizer Storage Mount.<br />

12. Remove the measuring equipment from the Common Gas Outlet.<br />

13. Reconnect the Fresh Gas, and Waste Gas Scavenger Hoses.<br />

6.8.10.3 Vaporizer Leak Test<br />

NOTE: Verify the system has passed the leak test without the<br />

vaporizers prior to this test.<br />

1. Verify that vaporizers are mounted correctly.<br />

2. Ensure that the gas pressure for O 2 , N 2 O, and AIR are at 50 ±10 psi.<br />

3. Power ON the AS3000.<br />

4. Attach a reusable-silicone-rubber breathing circuit to the Breathing System.<br />

NOTE: For testing purposes always use a reusable breathing<br />

circuit.<br />

5. Tightly connect the Y-fitting on the breathing circuit to the test port.<br />

6. Attach a breathing bag to the bag arm.<br />

7. Set the APL Valve to the fully closed position (70 cmH 2 O).<br />

8. Rotate the O 2 Flow Control Valve until 50 cmH 2O pressure is observed on the Airway<br />

Pressure Gauge.<br />

9. Verify that the necessary flow to stabilize the pressure is less than 300 mL/min.<br />

6 - 36 0070-10-0683 AS3000 Service Manual

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!