AS3000 Service Manual 0070-10-0683 6 - 3<br />
6.5 Replacement of Consumable Parts<br />
Parts are replaced at multiple intervals from the date of installation.<br />
CONSUMABLE PARTS AMOUNT 12 MONTH PM 36 MONTH PM PART NUMBER<br />
1 O 2 Sensor 1 check check 0600-00-0149<br />
2 Battery Assembly SLA-5AH24V 1 check replace 0997-00-0619<br />
3 Drive Gas Tube 1 check replace 0004-00-0098<br />
4 Bellows Assembly 1 replace replace 115-005968-00<br />
5 Scavenger Hose 1 check replace 0004-00-0095<br />
6 I/E Valve gasket 2 replace replace 0354-00-0192<br />
7 O-ring, 6 x 1.8 (for airway pressure gauge) 1 replace replace 0354-00-0188<br />
8 O-ring, 4 x 1.8 (for airway pressure gauge) 1 replace replace 0354-00-0189<br />
9 Bag Arm retainer ring 1 check check 0354-00-0039<br />
10 Disc, Ceramic I/E Valve 2 check replace 0354-00-0182<br />
11 Gasket, Bellows canister base 1 replace replace 0354-00-0191<br />
12 Absorber canister Seal<br />
3 check replace 0354-00-0183 (original)<br />
1<br />
2<br />
check<br />
check<br />
replace<br />
replace<br />
047-002850-00 (revised)<br />
047-000252-00 (revised)<br />
13 O-ring O2 Interface / O2 Plug 2 replace replace 0354-00-0187<br />
14 O-ring Vaporizer mount 14x2.65 4 replace replace 0354-00-0193<br />
15 Cylinder Washer, Black 3 replace replace 0348-00-0185<br />
16 Absorber assembly 1 clean clean 0997-00-0629 (original)<br />
115-005724-00 (revised)<br />
17 AGSS filter 1 clean replace 0378-00-0074<br />
18 Absorber Gasket 2 clean replace 0354-00-0199 (original)<br />
047-002849-00 (revised)<br />
19 Check Valve<br />
(Applicable to original model P/N 0997-00-0618<br />
only. See “Check Valve Cleaning” on page 6-4.)<br />
1 clean clean Non-replaceable<br />
20 User Interface cable jackscrews and jackposts 2 Check for Loctite Check for Loctite Non-replaceable<br />
243 and apply if 243 and apply if<br />
necessary. Verify necessary. Verify<br />
jackscrews are jackscrews are<br />
secured tightly. secured tightly.<br />
Periodic Maintenance Replacement of Consumable Parts