Theory of Operation Ventilator UI<br />

1.7.3 Communication Interface / RS232 Isolate Board<br />

1.7.4 Fuses<br />

FIGURE 1-26 RS232 Isolate Board, Top View<br />

FIGURE 1-27 RS232 Isolate Board, Bottom View<br />

RS232 Isolator<br />

Power, J3<br />

Pin Name Function<br />

4. 5V Power 5V<br />

5. NC No connection<br />

6. GND Ground<br />

COM1/2<br />

Pin Name Function<br />

2 RXD 232RXD TO PC INPUT<br />

3 TXD 232TXD TO PC OUTPUT<br />

5 GND Ground<br />

COM1A/B<br />

Pin Name Function<br />

2 TXD 232TXD TO PC OUTPUT<br />

3 RXD 232RXD TO PC INPUT<br />

5 GND Ground<br />

The AS3000 has 10 fuses on the back panel of the unit. There are two 2A fuses for each of<br />

the four outlets and two 10A fuses for the line cord.<br />