AS3000 Service Manual 0070-10-0683 3 - 7<br />

MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION<br />

1 Calibrate the inspiration valve again.<br />

In leakage detection, Safety<br />

valve detection failed.<br />

In leakage detection, the<br />

leakage volume is 501 -<br />

999mL.<br />

In leakage detection, the<br />

leakage volume is larger than<br />

1L.<br />

In compliance test,<br />

compliance test failed.<br />

The power supply switch on<br />

the main unit is defective.<br />

All or some keys are non<br />

responsive.<br />

The encoder is non<br />

responsive.<br />

All or some of the work lights<br />

are defective.<br />

The switch on the work light<br />

is defective.<br />

The flow meter light is non<br />

operational.<br />

Startup test 1 The proportional valve on the airway<br />

module failed.<br />

2 The safety valve in the airway module<br />

failed.<br />

Startup Test 1 The tubing connection is not tight.<br />

2 Incorrect installation of the absorber<br />

canister.<br />

3 The pressure gauge or oxygen sensor on the<br />

circuit is not installed tightly.<br />

4 There is a leakage in the fresh gas supply.<br />

Startup Test 1 The tubing connection is not tight.<br />

2 Incorrect installation of the absorber<br />

canister.<br />

3 The pressure gauge or oxygen sensor on the<br />

circuit is not installed tightly.<br />

4 There is a leakage in the fresh gas supply.<br />

Startup Test 1 The Y-fitting is not sealed or the Breathing<br />

System has leakage.<br />

2 The respiration tubes are too long.<br />

3 Calibration of the pressure sensor is<br />

inaccurate.<br />

4 The flow meter is set incorrectly<br />

5 The calibration of the flow meter is<br />

inaccurate<br />

All 1 The power supply switch on the main unit or<br />

the cable connection failed.<br />

2 Replace the airway module.<br />

1 Check the connection on the tubing.<br />

2 Check the installation of the absorber canister.<br />

3 Check the installation of the pressure gauge or<br />

oxygen sensor on circuit.<br />

4 Check the fresh gas supply for leakage<br />

1 Check the connection on the tubing.<br />

2 Check the installation of the absorber canister.<br />

3 Check the installation of the pressure gauge or<br />

oxygen sensor on circuit.<br />

4 Check the fresh gas supply for leakage<br />

1 Check the connection and leakage of<br />

respiration tubes.<br />

2 The respiration tubes cannot be too long.<br />

3 Calibrate the pressure oxygen again.<br />

4 Set the flow meter to the correct value<br />

5 Recalibrate the flow meter<br />

1 Check the connection cable of the power<br />

supply switch of the main unit, or replace the<br />

power supply switch of main unit.<br />

2 Communication board is defective.<br />

2 Replace the communication board.<br />

All 1 The keyboard is defective.<br />

1 Replace the user interface assembly.<br />

2 The communication board is defective. 2 Replace the user interface assembly.<br />

All 1 The encoder failed.<br />

1 Replace the user interface assembly.<br />

2 The communication board is defective. 2 Check and replace the user interface assembly.<br />

All 1 The work light circuit board is defective. 1 Replace the work light circuit board.<br />

2 The cable is defective<br />

2 Check the connection to the power supply.<br />

All 1 The switch on the work light is defective. 1 Replace the work light switch.<br />

All 1 The power supply to the light is defective<br />

2 The foil of light is defective.<br />

3 The inverter is defective.<br />

1 Replace the power supply to the light.<br />

2 Replace the foil of light.<br />

3 Replace the inverter<br />

Repair Information Troubleshooting Chart