AS3000 Service Manual 0070-10-0683 3 - 7<br />
MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION<br />
1 Calibrate the inspiration valve again.<br />
In leakage detection, Safety<br />
valve detection failed.<br />
In leakage detection, the<br />
leakage volume is 501 -<br />
999mL.<br />
In leakage detection, the<br />
leakage volume is larger than<br />
1L.<br />
In compliance test,<br />
compliance test failed.<br />
The power supply switch on<br />
the main unit is defective.<br />
All or some keys are non<br />
responsive.<br />
The encoder is non<br />
responsive.<br />
All or some of the work lights<br />
are defective.<br />
The switch on the work light<br />
is defective.<br />
The flow meter light is non<br />
operational.<br />
Startup test 1 The proportional valve on the airway<br />
module failed.<br />
2 The safety valve in the airway module<br />
failed.<br />
Startup Test 1 The tubing connection is not tight.<br />
2 Incorrect installation of the absorber<br />
canister.<br />
3 The pressure gauge or oxygen sensor on the<br />
circuit is not installed tightly.<br />
4 There is a leakage in the fresh gas supply.<br />
Startup Test 1 The tubing connection is not tight.<br />
2 Incorrect installation of the absorber<br />
canister.<br />
3 The pressure gauge or oxygen sensor on the<br />
circuit is not installed tightly.<br />
4 There is a leakage in the fresh gas supply.<br />
Startup Test 1 The Y-fitting is not sealed or the Breathing<br />
System has leakage.<br />
2 The respiration tubes are too long.<br />
3 Calibration of the pressure sensor is<br />
inaccurate.<br />
4 The flow meter is set incorrectly<br />
5 The calibration of the flow meter is<br />
inaccurate<br />
All 1 The power supply switch on the main unit or<br />
the cable connection failed.<br />
2 Replace the airway module.<br />
1 Check the connection on the tubing.<br />
2 Check the installation of the absorber canister.<br />
3 Check the installation of the pressure gauge or<br />
oxygen sensor on circuit.<br />
4 Check the fresh gas supply for leakage<br />
1 Check the connection on the tubing.<br />
2 Check the installation of the absorber canister.<br />
3 Check the installation of the pressure gauge or<br />
oxygen sensor on circuit.<br />
4 Check the fresh gas supply for leakage<br />
1 Check the connection and leakage of<br />
respiration tubes.<br />
2 The respiration tubes cannot be too long.<br />
3 Calibrate the pressure oxygen again.<br />
4 Set the flow meter to the correct value<br />
5 Recalibrate the flow meter<br />
1 Check the connection cable of the power<br />
supply switch of the main unit, or replace the<br />
power supply switch of main unit.<br />
2 Communication board is defective.<br />
2 Replace the communication board.<br />
All 1 The keyboard is defective.<br />
1 Replace the user interface assembly.<br />
2 The communication board is defective. 2 Replace the user interface assembly.<br />
All 1 The encoder failed.<br />
1 Replace the user interface assembly.<br />
2 The communication board is defective. 2 Check and replace the user interface assembly.<br />
All 1 The work light circuit board is defective. 1 Replace the work light circuit board.<br />
2 The cable is defective<br />
2 Check the connection to the power supply.<br />
All 1 The switch on the work light is defective. 1 Replace the work light switch.<br />
All 1 The power supply to the light is defective<br />
2 The foil of light is defective.<br />
3 The inverter is defective.<br />
1 Replace the power supply to the light.<br />
2 Replace the foil of light.<br />
3 Replace the inverter<br />
Repair Information Troubleshooting Chart