Periodic Maintenance Check Valve Cleaning<br />

6.6.2 Cleaning Procedure<br />

1. Remove Absorber Locking Mechanism and Absorber Mount Assembly:<br />

1. Rotate the cam handle to the unlocked position.<br />

2. Remove (2) Absorber Canister Assemblies.<br />

3. Remove Absorber Hose from port on Absorber Mount Assembly.<br />

4. Remove (2) 4mm hex screws securing the Absorber Locking Mechanism to the rods.<br />

5. Remove the Absorber Locking Mechanism and Absorber Mount Assembly.<br />

Absorber Mount<br />

Assembly<br />

Absorber Canister<br />

Assembly<br />

FIGURE 6-2 Absorber Assembly<br />

Absorber<br />

Locking<br />

Mechanism<br />

Absorber<br />

Hose<br />

AS3000 Service Manual 0070-10-0683 6 - 5