Microprocessor-controlled Ventilator Theory of Operation<br />
1.2 Microprocessor-controlled Ventilator<br />
The Microprocessor-controlled ventilator, with its dedicated Breathing System, allows time-<br />
controlled, pressure limited, constant volume ventilation for all patient groups within a tidal<br />
volume range of 40 mL (4 kg infant) to 1400 mL (large adult).<br />
Time-controlled, pressure limited, and compliance compensated constant volume ventilation<br />
is provided through the Controlled Mandatory Ventilation (CMV) mode. The CMV mode<br />
delivers a viable ventilation method for complicated lung conditions. The ventilator also<br />
provides time-controlled, volume dependant ventilation, targeting a set (adjustable) target<br />
pressure provided through the Pressure Controlled Ventilation (PCV) mode. Automatic and<br />
comprehensive system startup tests and alarm management systems ensure controlled<br />
ventilation conditions in every mode of operation.<br />
The durable and ergonomically designed user interface and Navigator Knob enables easy<br />
operation. The display provides the selected ventilation modes (CMV, PCV, PS and SIMV)<br />
and the following values: Tidal Volume, Peak Pressure, Mean Pressure, FiO2, Breath Rate, I:E<br />
Ratio, PEEP, Plateau Pressure, Alarm limits, and real-time Airway Pressure and flow<br />
waveform.<br />
Patient<br />
Electronic<br />
Control<br />
Exp. Flow<br />
Sensor<br />
Insp. Flow<br />
Sensor<br />
P<br />
Exp. Valve<br />
O2 Sensor<br />
Insp. Valve<br />
Absorber<br />
Canister<br />
Fresh Gas<br />
Breathing Bag<br />
Solenoid<br />
Valve<br />
Reversing<br />
Valve<br />
Vaporizer<br />
Proportional<br />
Valve<br />
Pressure<br />
Regulator<br />
FIGURE 1-1 The AS3000 Pneumatic System<br />
O2<br />
Flush<br />
APL Valve<br />
Bellows<br />
Vaporizer<br />
Exhaust<br />
Valve<br />
PEEP<br />
Valve<br />
Fresh Gas<br />
Flow Sensor<br />
V<br />
Breathing<br />
System<br />
Ventilator<br />
Ratio<br />
Controller<br />
N20<br />
Cuttoff<br />
Gas Inlet<br />
Block<br />
1 - 2 0070-10-0683 AS3000 Service Manual<br />
AGSS<br />
Flow<br />
Tubes<br />
AIR<br />
Flow<br />
Meter<br />
Pressure<br />
Regulator<br />
Pressure<br />
Regulator<br />
Pressure<br />
Regulator<br />
O2<br />
Flow<br />
Meter<br />
O2<br />
Outlet<br />
Pressure<br />
Switch<br />
Inlet Filter<br />
Inlet Filter<br />
Inlet Filter<br />
P<br />
P<br />
P<br />
P<br />
P<br />
P<br />
Cylinder<br />
O O O<br />
Cylinder<br />
Cylinder<br />
N20 AIR O2