Check Valve Cleaning Periodic Maintenance<br />

6.6 Check Valve Cleaning<br />

Check valve cleaning is applicable only the original beathing system model (P/N 0997-00-<br />

0618) only. It does not need to be performed on the revised beathing system model (P/N<br />

115-005967-00) (see FIGURE 6-1).<br />

Constant contact with moist patient gas can result in the breathing system check valve<br />

becoming sticky. Cleaning the check valve will resolve this issue. Perform this cleaning<br />

procedure at least every 12 months.<br />

NOTE: Plastic gloves should be worn when handling breathing<br />

systems<br />

6.6.1 Tools and Materials<br />

New Style Breathing Systems<br />

(115-005967-00)<br />

FIGURE 6-1 Breathing System Styles<br />

Philips screw driver<br />

2.5 mm hex wrench<br />

3 mm hex wrench<br />

4 mm hex wrench<br />

12 mm standard socket<br />

13 mm open end wrench<br />

Silicone Lubricant (PN 0510-00-0021)<br />

Tape measure (metric)<br />

Loctite 243<br />

Isopropyl alcohol<br />

Old Style Breathing Systems<br />

(0997-00-0618)<br />

6 - 4 0070-10-0683 AS3000 Service Manual