Ventilator Control and Drive Theory of Operation<br />
Signals and Power to Keyboard, J5 (Continued)<br />
PIN NAME FUNCTION<br />
10. PWBUS PowerBus<br />
11. PWBUS PowerBus<br />
12. PWBUS PowerBus<br />
13. GND Ground<br />
14. GND Ground<br />
15. DC24V DC24V monitor<br />
16. BATV+ Battery monitor<br />
17. /MUTE No connection<br />
18. GND Ground<br />
19. TTLTX TTL Transmit<br />
20. 232ATX RS232A Transmit<br />
21. 232BTX RS232B Transmit<br />
22. GND Ground<br />
23. PWBUS PowerBus<br />
24. PWBUS PowerBus<br />
25. GND Ground<br />
Signals and Power to Sensor Board, J30<br />
PIN NAME FUNCTION<br />
1. A+12V Analogue +12V<br />
2. FL_INS Import flow<br />
3. FL_EXP Export flow<br />
4. FM_FL Flowmeter flow<br />
5. Pair Pressure of air<br />
6. Paw Pressure inside air way<br />
7. Pgas Pressure gas supply status<br />
8. GND Ground<br />
9. AGND Analog Ground<br />
10. AGND Analog Ground<br />
11. P24V Power 24V<br />
12. P24V Power 24V<br />
13. P24V Power 24V<br />
14. P12V Power 12V<br />
15. P12V Power 12<br />
16. P5V Power 5V<br />
17. P5V Power 5V<br />
18. Inhale Inspire Valve<br />
19. Man Manual/AUTO Valve<br />
20. Exhale Expire Valve<br />
21. Exhale Expire Valve<br />
22. AGND Analog Ground<br />
1 - 16 0070-10-0683 AS3000 Service Manual