Theory of Operation Ventilator Control and Drive<br />

Power Board<br />

Battery input, J4<br />

PIN NAME FUNCTION<br />

1. BAT Battery Input +<br />

2. 5V Power Ground<br />

3. BATTEST Battery test input<br />

DC24V Power input, J3<br />

PIN NAME FUNCTION<br />

4. GND Power Ground<br />

5. GND Power Ground<br />

6. 24V Power input<br />

7. 24V Power input<br />

DC Power output, J2<br />

PIN NAME FUNCTION<br />

1. GND Power Ground<br />

2. +5V Controller Logic Power<br />

3. P12V Power 12V<br />

4. PWBUS Power Bus 24V<br />

5. GND Analog Ground<br />

6. A12V Analog 12V<br />

7. 5V/4.8V Power 5V/4.8V<br />

Heater Power, J1<br />

PIN NAME FUNCTION<br />

6. GND Power Ground<br />

7. 5V Controller Power, Output Enable<br />

8. 24V Heater Power Supply<br />

9. PBUS Power Bus, No Connection<br />

Power Control, J2<br />

PIN NAME FUNCTION<br />

1. GND Ground<br />

2. /PWON Power On Enable, Low active<br />

3. PWEN Power On Enable, High active<br />

4. 5VAUX 5V Auxiliary<br />

5. DC_M AC/DC Monitor signal<br />

6. PWBUS Power bus<br />

7. BAT_M Battery Monitor signal<br />

AS3000 Service Manual 0070-10-0683 1 - 27