Periodic Maintenance Preoperative Checklist<br />

16. Test the Ventilator in STANDBY mode:<br />

a. Set the APL Valve to 20 cmH 2 O.<br />

b. Attach a test lung to the breathing circuit Y-fitting.<br />

c. Set the AIR flow to 5 L/min.<br />

d. Press the O 2 Flush Valve to fill the breathing bag with volume.<br />

e. Squeeze the breathing bag once every 10 seconds.<br />

f. Verify that the test lung inflates and deflates.<br />

g. Set the APL Valve to the open position (SP) and stop squeezing the breathing bag.<br />

17. Test the Ventilator in MANUAL mode:<br />

a. Set the ventilation mode to MANUAL.<br />

b. Set the APL Valve to 20 cmH 2O.<br />

c. Set the AIR flow to 5 L/min.<br />

d. Squeeze the breathing bag once every 10 seconds.<br />

e. Verify that the test lung inflates and deflates to approximately 20 cmH 2O pressure.<br />

f. Verify that tidal volume, resp. rate, minute volume, and PEEP values appear on the<br />

screen.<br />

g. Verify that the FiO 2 reading is approximately 21.<br />

h. Verify a pressure waveform appears on the screen along with the bag compressions.<br />

i. Set the APL Valve to the open position (SP) and stop squeezing the breathing bag.<br />

18. Test the Tidal Volume in CMV mode:<br />

a. Set the O 2 flow to 2 L/min and set the N 2O and AIR flow rates to minimum flow.<br />

b. Set the ventilator controls to:<br />

ATTRIBUTE SETTING<br />

Patient Type Adult<br />

Ventilation Mode CMV<br />

Tidal Volume - VT 600<br />

Breathing Rate - freq 8<br />

I:E Ratio - I:E 1:2<br />

Plateau - TP 10<br />

PEEP - PEEP OFF<br />

c. Select CMV to begin ventilation.<br />

d. Verify that the Tidal Volume display is within 15% of the set value within 5 breaths.<br />

e. Verify that the O 2 display reads greater than 95% within 5 minutes.<br />

AS3000 Service Manual 0070-10-0683 6 - 43