Installation Guide Assembly<br />
2. Discard the washers that came with the vaporizer. Use only approved O-rings that come<br />
mounted on the vaporizer mounting ports of the AS3000.<br />
3. Mount a mechanical vaporizer to either side of the AS3000 vaporizer mount by<br />
placing the vaporizer on the ports and locking down the lever.<br />
4. Mount an electronic vaporizer in the same manner. Follow any installation instructions<br />
that come with the specific vaporizer.<br />
2.2.8 High Pressure Hoses<br />
1. Remove the hoses from their packing material.<br />
2. Attach each hose to its mating connector by hand. Do not use any tools.<br />
3. Use extreme care while attaching the EVAC hose to the waste gas scavenger (WGS); as<br />
it is extremely delicate.<br />
4. Attach an EVAC hose to a VAC source connection only if an EVAC source connection is<br />
not available.<br />
2.2.9 Emergency Cylinder(s)<br />
1. Remove the cover from a new O 2, N 2O, and AIR cylinder.<br />
2. Mount one at a time onto the rear of the anesthesia machine.<br />
3. Discard the cylinder's tank washer. Always use the approved tank washer provided with<br />
the AS3000.<br />
4. Open the bail of each yoke and mount the cylinder over the tank washer.<br />
5. Ensure the O 2 cylinder mates to the O 2 Pin Index Safety System (PISS) connection on the<br />
O 2 yoke. Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke<br />
port.<br />
6. Open and close the cylinder valve and observe that the cylinder gauge on the<br />
anesthesia machine rises to the colored range.<br />
7. Ensure that the N 2O cylinder mates to the N 2O PISS connection on the N 2O yoke.<br />
Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke port.<br />
8. Open and close the cylinder valve and observe that the cylinder gauge on the<br />
anesthesia machine rises to the colored range.<br />
9. Ensure that the AIR cylinder mates to the AIR PISS connection on the AIR yoke. Close the<br />
yoke bail and use the hand-screw to tighten the cylinder to the yoke port.<br />
10. Open and close the cylinder valve and observe the cylinder gauge on the anesthesia<br />
machine rise to the colored range.<br />
2.2.10 Breathing Circuit, CO 2 Absorbent, and Liquid Vaporizer Agent<br />
1. Attach a breathing circuit to the inspiratory and expiratory ports as detailed in the<br />
directions for use. Attach the breathing bag and any other respiratory accessories as<br />
described.<br />
2. Insert the Pre-Paks one on top of the other with the wider side facing up on both Pre-<br />
Paks. (alternatively the absorber canisters may be filled with loose fill absorbent.)<br />
3. Install the absorber canister with a quarter turn of the lever at the bottom the absorber<br />
assembly, this ensures a tight seal.<br />
AS3000 Service Manual 0070-10-0683 2 - 5