2.0 Installation Guide<br />

2.1 Delivery of The New Anesthesia Machine<br />

The following customer supplied material must be present prior to installation. Missing<br />

equipment can result in delays, incomplete installations and/or extra visits.<br />

Compatible emergency O 2, N 2O, and AIR cylinders<br />

Agent vaporizers and key fillers, if not purchased with the AS3000<br />

Liquid agent medication<br />

CO2 absorbent Pre-Paks or loose fill<br />

Active O2, N2O, and AIR, lines at 50 psi<br />

Drop down hoses for ceiling mounted medical gas utilities, compatible with quickdisconnect<br />

hoses if not purchased with the AS3000<br />

Activated medical gases (O2 , N2O, AIR, VAC, and EVAC)<br />

AS3000 Service Manual 0070-10-0683 2 - 1