Functional Tests Periodic Maintenance<br />

6.8.8 Alarms and Failsafe Functions<br />

6.8.8.1 Set Up<br />

1. Ensure that the gas pressure for O 2 , N 2 O, and AIR are at 50 ±10 psi.<br />

2. Power ON the AS3000.<br />

3. Perform the Startup Tests per the on-screen instructions. Ensure successful completion.<br />

4. Attach a breathing circuit and breathing bag.<br />

NOTE: For testing purposes always use a reusable breathing<br />

circuit.<br />

5. Attach an adult test lung to the Y-fitting of the breathing circuit.<br />

6. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.<br />

7. Set the ventilator controls to:<br />

VENTILATOR CONTROLS VENTILATOR SETTINGS<br />

Patient Type Adult<br />

Ventilation Mode CMV<br />

Tidal Volume - VT 600<br />

Breath Rate - freq 8<br />

I:E Ratio - I:E 1:2<br />

Plateau - TP 10<br />

PEEP - PEEP Off<br />

8. Select CMV to begin ventilation.<br />

6.8.8.2 Low FiO 2 Alarm Test<br />

1. Set the low FiO 2 Alarm limit to 50% O 2.<br />

2. Set the AIR flow control valve to 5 L/min.<br />

3. Set the FiO2 flow controller to minimum flow.<br />

4. Verify the following Low FiO2 alarm signals activate, within three ventilation cycles:<br />

Low FiO2 message appears on the screen.<br />

An alarm tone sounds.<br />

5. Set the Low FiO2 alarm limit to 18%.<br />

6. Verify the alarm signals cease.<br />

6 - 32 0070-10-0683 AS3000 Service Manual