28.02.2013 Views

Datascope - Mindray

Datascope - Mindray

Datascope - Mindray

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

AS3000 Service Manual 0070-10-0683 3 - 5<br />

MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION<br />

“Vent/Manual Valve Failure” Startup test 1 The wire that connects to the vent/manual<br />

valve is disconnected or is damaged.<br />

2 The vent/manual solenoid valve is<br />

damaged.<br />

3 The amplifier board or connection cable<br />

defective.<br />

“Expiration Sensor Failure” Startup test 1 Fresh gas flow is present at startup.<br />

“Keyboard Communication .<br />

. . FAIL”<br />

2 The sensor board is defective.<br />

Startup test 1 The data wire is disconnected or is<br />

damaged.<br />

2 The keyboard board is defective.<br />

“Pressure Sensor . . . FAIL” Startup test 1 The connection wire of the PAW board is<br />

disconnected or is damaged.<br />

2 The connection of the PAW board or sensor<br />

board is defective.<br />

3 The tubing kinked or occluded.<br />

“O 2 Sensor . . . FAIL” Startup test 1 The O 2 concentration is below 16%.<br />

2 The oxygen sensor is out of range.<br />

3 The main unit-oxygen sensor cable is<br />

disconnected or damaged.<br />

4 The amplifier board is defective.<br />

“PEEP Valve . . . FAIL” Startup test 1 The connection wire of the PEEP valve is<br />

disconnected or damaged.<br />

2 The PEEP valve is defective.<br />

3 The Amplifier board is defective.<br />

“Inspiration Valve . . . FAIL” Startup test 1 The connection wire of the inspiration valve<br />

is disconnected or damaged.<br />

2 The inspiration valve is defective.<br />

3 The amplifier board is defective.<br />

1 Check whether the connection wire of the<br />

vent/manual valve is disconnected or is<br />

damaged.<br />

2 Replace the vent/manual solenoid valve.<br />

3 Check the connection cable and replace the<br />

amplifier board. Then perform each item of the<br />

calibration and startup tests again.<br />

1 Turn off all fresh gas flow and reboot the unit.<br />

2 Replace the sensor board, and then perform<br />

each item of the calibration and startup test<br />

again.<br />

1 Replace the user interface assembly.<br />

2 Replace the user interface assembly.<br />

1 Check whether the connection wire of the PAW<br />

board is disconnected or damaged.<br />

2 Check the power supply and connection of the<br />

sensor board. Replace the Paw board and<br />

sensor board.<br />

3 Check the tubing for kinks and remove any<br />

occlusions.<br />

1 Increase O2 concentration.<br />

2 Replace the oxygen sensor.<br />

3 Check whether the main unit-oxygen sensor<br />

cable is disconnected or damaged.<br />

4 Replace the amplifier board, and then perform<br />

each item of the calibration and startup test<br />

again.<br />

1 Check whether the PEEP valve is disconnected<br />

or damaged.<br />

2 Replace the PEEP valve.<br />

3 Replace the amplifier board and then perform<br />

each item of the calibration and startup test<br />

again.<br />

1 Check whether the inspiration valve is<br />

disconnected or damaged.<br />

2 Replace the inspiration valve.<br />

3 Replace the amplifier board and then perform<br />

each item of the calibration and startup test<br />

again.<br />

Repair Information Troubleshooting Chart

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!