Components Theory of Operation<br />

26. Vaporizer Storage Mount<br />

The Vaporizer Storage Mount provides the AS3000 with auxiliary placement for a<br />

vaporizer.<br />

27. Mains Supply Assembly<br />

The Mains Supply Assembly attaches the AS3000 to an external AC wall outlet. It also<br />

supplies power throughout the unit.<br />

28. Electronic Flowmeter Assembly<br />

The Electronic Flowmeter Assembly provides fresh gas flow measurement for the AS3000.<br />

29. Gas Box Assembly<br />

The Gas Box Assembly provides drive gas to the Breathing System and includes a safety<br />

valve to insure that pressure stays below 85 cmH2O. It also contains the PEEP valve, and the<br />

differential pressure sensors.<br />

30. YOKE Assembly<br />

The YOKE Assembly provides connection for three backup gas cylinders. The pressure<br />

regulator for the backup gas cylinder reduces the pressure of a high pressure gas cylinder<br />

down to 43.5 psi (300 kPa).<br />

31. AMP Cables<br />

The AMP Cables provide signal transmission between the electric box and gas circuit box.<br />

32. Electric Box Assembly<br />

The Electric Box Assembly provides, and distributes power to the AS3000.<br />

33. Rear Panel Assembly<br />

The Rear Panel Assembly protects, and provides heat dissipation for the electric box and<br />

auxiliary support for the backup gas cylinder.<br />

1.3.4 Other components (not identified in graphics)<br />

34. Inspiratory and Expiratory Valves<br />

The Inspiratory and Expiratory Valves are unidirectional valves that allow air to flow in only<br />

one direction.<br />

35. Absorber Canister<br />

The AS3000 uses two Absorber Canisters which can contain 1500 mL of soda lime each.<br />

They can be used for 6 - 8 hours each if full. Water generated from the reaction with CO2 is<br />

drained from a valve on the lower side of the canister.<br />

36. APL (Airway Pressure Limiting) Valve<br />

The APL Valve is used for limiting maximum airway pressure during manual ventilation. The<br />

adjustable range of the APL is 0 - 70 cmH2O. 1 - 8 0070-10-0683 AS3000 Service Manual