Calibration Calibration Procedures<br />

5.5.4 Flow Valve Calibration<br />

1. Preparing the unit<br />

FIGURE 5-5 Connecting the breathing circuit<br />

a. Reinstall the Inspiratory and Expiratory valve disks.<br />

b. Connect a breathing circuit to the Breathing System (see FIGURE 5-5).<br />

c. Connect a 3L bag to the Y-fitting.<br />

NOTE: A 2.3L bag can be used in place of a 3L bag.<br />

d. Select Continue.<br />

2. Calibration<br />

a. Press the O 2 Flush Valve to fully fill bellows.<br />

b. Select Continue to start calibration.<br />

c. Follow the on-screen instructions.<br />

NOTE: The flow Valve calibration duration is approximately 5<br />

minutes.<br />

AS3000 Service Manual 0070-10-0683 5 - 9