Calibration Calibration Procedures<br />
5.5.4 Flow Valve Calibration<br />
1. Preparing the unit<br />
FIGURE 5-5 Connecting the breathing circuit<br />
a. Reinstall the Inspiratory and Expiratory valve disks.<br />
b. Connect a breathing circuit to the Breathing System (see FIGURE 5-5).<br />
c. Connect a 3L bag to the Y-fitting.<br />
NOTE: A 2.3L bag can be used in place of a 3L bag.<br />
d. Select Continue.<br />
2. Calibration<br />
a. Press the O 2 Flush Valve to fully fill bellows.<br />
b. Select Continue to start calibration.<br />
c. Follow the on-screen instructions.<br />
NOTE: The flow Valve calibration duration is approximately 5<br />
minutes.<br />
AS3000 Service Manual 0070-10-0683 5 - 9