Installation Checkout Procedure Installation Guide<br />

e. Set the ventilator controls to:<br />

VENTILATOR CONTROLS VENTILATOR SETTINGS<br />

Patient Type Adult<br />

Ventilation Mode CMV<br />

Tidal Volume - VT 600<br />

Breath Rate - freq 8<br />

I:E Ratio - I:E 1:2<br />

Plateau - TP 10<br />

PEEP - PEEP Off<br />

f. Select CMV again to begin ventilation.<br />

g. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp.<br />

rate and minute volume values appear on the screen.<br />

h. Verify the Tidal Volume display is within 15% of the delivered volume measured with<br />

the Vent Tester within approximately 1 minute from the start of ventilation.<br />

i. Verify the delivered volume at the Vent Tester is within 15% of the Tidal Volume set<br />

value within approximately 1 minute from the start of ventilation.<br />

j. Verify the measured O 2 concentration is at least 97% after 5 minutes.<br />

k. Set the AIR flow to 3 L/min and set the N 2O and O 2 flow rates to minimum flow.<br />

l. Verify the measured O 2 concentration is 21% ±3% vol. % after 5 minutes.<br />

19. Verify that CMV CHILD ventilator mode operates and that the tidal<br />

volume display reads within 10% of the CMV set value.<br />

a. Attach a breathing circuit and breathing bag.<br />

NOTE: For testing purposes always use a reusable breathing<br />

circuit.<br />

b. Attach an adult test lung to the Y-fitting of the breathing circuit.<br />

NOTE: Limit the volume in the test lung to provide sufficient airway<br />

pressure to satisfy the Low Peak Pressure alarm. Or reduce<br />

the Peak Pressure alarm limit to a lower valve to prevent<br />

the alarm when using an adult test lung.<br />

2 - 20 0070-10-0683 AS3000 Service Manual