Installation Checkout Procedure Installation Guide<br />
e. Set the ventilator controls to:<br />
VENTILATOR CONTROLS VENTILATOR SETTINGS<br />
Patient Type Adult<br />
Ventilation Mode CMV<br />
Tidal Volume - VT 600<br />
Breath Rate - freq 8<br />
I:E Ratio - I:E 1:2<br />
Plateau - TP 10<br />
PEEP - PEEP Off<br />
f. Select CMV again to begin ventilation.<br />
g. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp.<br />
rate and minute volume values appear on the screen.<br />
h. Verify the Tidal Volume display is within 15% of the delivered volume measured with<br />
the Vent Tester within approximately 1 minute from the start of ventilation.<br />
i. Verify the delivered volume at the Vent Tester is within 15% of the Tidal Volume set<br />
value within approximately 1 minute from the start of ventilation.<br />
j. Verify the measured O 2 concentration is at least 97% after 5 minutes.<br />
k. Set the AIR flow to 3 L/min and set the N 2O and O 2 flow rates to minimum flow.<br />
l. Verify the measured O 2 concentration is 21% ±3% vol. % after 5 minutes.<br />
19. Verify that CMV CHILD ventilator mode operates and that the tidal<br />
volume display reads within 10% of the CMV set value.<br />
a. Attach a breathing circuit and breathing bag.<br />
NOTE: For testing purposes always use a reusable breathing<br />
circuit.<br />
b. Attach an adult test lung to the Y-fitting of the breathing circuit.<br />
NOTE: Limit the volume in the test lung to provide sufficient airway<br />
pressure to satisfy the Low Peak Pressure alarm. Or reduce<br />
the Peak Pressure alarm limit to a lower valve to prevent<br />
the alarm when using an adult test lung.<br />
2 - 20 0070-10-0683 AS3000 Service Manual