The Breathing System Theory of Operation<br />
1.9.3 Manual mode, inspiration<br />
As the breathing bag is compressed, the gas is directed to the patient. Pressures exceeding<br />
the set value of the APL Valve will pass through the APL Valve the AGSS.<br />
Patient<br />
FIGURE 1-30 Breathing System Pneumatics, Manual Mode, Inspiration<br />
1.9.4 Manual mode, expiration<br />
As the patient exhales tidal volume into the expiratory limb, fresh gas enters the Breathing<br />
System. Fresh gas mixes with exhaled gas after the Absorber removes CO 2. Excess fresh gas<br />
passes through the exhaust valve to the AGSS<br />
Patient<br />
Electronic<br />
Control<br />
Electronic<br />
Control<br />
Exp. Flow<br />
Sensor<br />
Insp. Flow<br />
Sensor<br />
Exp. Flow<br />
Sensor<br />
Insp. Flow<br />
Sensor<br />
P<br />
O2 Sensor<br />
Insp. Valve<br />
P<br />
Exp. Valve<br />
Exp. Valve<br />
O2 Sensor<br />
Insp. Valve<br />
Absorber<br />
Canister<br />
Absorber<br />
Canister<br />
Fresh Gas<br />
Fresh Gas<br />
Solenoid<br />
Valve<br />
Solenoid<br />
Valve<br />
Reversing<br />
Valve<br />
FIGURE 1-31 Breathing System Pneumatics, Manual Mode, Expiration<br />
Breathing Bag<br />
Breathing Bag<br />
Reversing<br />
Valve<br />
Breathing<br />
System<br />
Ventilator<br />
1 - 38 0070-10-0683 AS3000 Service Manual<br />
Proportional<br />
Valve<br />
Pressure<br />
Regulator<br />
Proportional<br />
Valve<br />
Pressure<br />
Regulator<br />
APL Valve<br />
Bellows<br />
APL Valve<br />
Bellows<br />
Exhaust<br />
Valve<br />
2 cmH2O<br />
PEEP<br />
Valve<br />
Exhaust<br />
Valve<br />
2 cm H2O<br />
PEEP<br />
Valve<br />
AGSS<br />
AGSS<br />
Breathing<br />
System<br />
Ventilator