Installation Checkout Procedure Installation Guide<br />

NOTE: If the system is going to be used during the calibration,<br />

insert the O 2 cell plug into the port from which the oxygen<br />

sensor was removed using a push and turn motion.<br />

e. After at least 3 minutes have passed, select the Next button to initiate the<br />

calibration process. The progress bar shown in FIGURE 2-15 or FIGURE 2-17will be<br />

displayed.<br />

FIGURE 2-14 Oxygen Sensor<br />

Calibration Instructions<br />

FIGURE 2-16 Oxygen Sensor<br />

Calibration Instructions (UI<br />

versions 2.24 and lower)<br />

FIGURE 2-15 Oxygen Sensor Calibration<br />

Progress Bar<br />

FIGURE 2-17 Oxygen Sensor Calibration<br />

Progress Bar (UI versions<br />

2.24 and lower)<br />

2 - 18 0070-10-0683 AS3000 Service Manual