Installation Checkout Procedure Installation Guide<br />
NOTE: If the system is going to be used during the calibration,<br />
insert the O 2 cell plug into the port from which the oxygen<br />
sensor was removed using a push and turn motion.<br />
e. After at least 3 minutes have passed, select the Next button to initiate the<br />
calibration process. The progress bar shown in FIGURE 2-15 or FIGURE 2-17will be<br />
displayed.<br />
FIGURE 2-14 Oxygen Sensor<br />
Calibration Instructions<br />
FIGURE 2-16 Oxygen Sensor<br />
Calibration Instructions (UI<br />
versions 2.24 and lower)<br />
FIGURE 2-15 Oxygen Sensor Calibration<br />
Progress Bar<br />
FIGURE 2-17 Oxygen Sensor Calibration<br />
Progress Bar (UI versions<br />
2.24 and lower)<br />
2 - 18 0070-10-0683 AS3000 Service Manual