Periodic Maintenance Battery Maintenance and Replacement<br />

6.7 Battery Maintenance and Replacement<br />

6.7.1 Battery Maintenance<br />

1. The AS3000 uses a sealed lead-acid battery. Due to the self-discharge characteristics<br />

of this type of battery, it is imperative that it is charged after 3 months of storage (or after<br />

extended periods of non-use). If not charged, a permanent loss of capacity may occur as<br />

a result of sulfation.<br />

2. Check the battery run time every 12 months.<br />

3. Replace batteries when operating time is less than 30 minutes or when the battery is 3<br />

years old.<br />

NOTE: The sealed lead-acid batteries used in the system are<br />

maintenance free. They will perform reliably provided that<br />

they are kept in the charged state.<br />

4. Disposal of batteries should be conducted in accordance with local recycling statues<br />

and labeling shown on the battery pack.<br />

6.7.2 Battery Replacement<br />

1. Open the access door located at the rear of the AS3000.<br />

2. Disconnect the battery connections.<br />

3. Remove the old battery.<br />

4. Ensure the new battery is electrically connected and secured in the same manner as the<br />

original one.<br />

5. Close the access door.<br />

6. Use only Approved Batteries (P/N 0997-00-0619).<br />

6.8 Functional Tests<br />

Refer to “Calibration” on page 5-1 if any values are out of specification.<br />

6.8.1 Test Equipment and Special Tools Required<br />

Refer to “Special Tools Required” on page 3-3 for a list of tools required for Running the<br />

functional tests.<br />

AS3000 Service Manual 0070-10-0683 6 - 13