Periodic Maintenance Functional Tests<br />

6.8.10.4 Dräger Vapor 2000 Operating Instructions ARRB-F001<br />

1. Fill Vaporizer - at least half full between minimum and maximum mark.<br />

2. Allow the filled Vapor to warm up to room temperature of 20-24�C. Wait long enough<br />

for the temperature to equalize - the time will vary depending on the temperature<br />

differential '�T'.<br />

1 hour = up top 2�C<br />

3 hours = ±6�C<br />

4 hours = ±10�C<br />

5 hours = ±20�C<br />

3. Check anesthetic agent monitor. Perform zero calibration of monitor with the desired<br />

gas (AIR or O 2 )<br />

4. Connect monitor to fresh gas outlet or Y-fitting. Make sure that all connections are leak-tight.<br />

5. Connect and start scavenging system.<br />

6. Switch OFF ventilator or set vent pressure to less than 5 cmH 2 O.<br />

7. Set monitor to anesthetic agent being used and to continuous measurement.<br />

8. Set flow between 2.5 and 4 L/min AIR. Use O 2 if AIR is not available.<br />

9. Check 0 and T marks, 1 vol.% 4 vol % and at least three other concentrations.<br />

10. Adjust control dial on the vaporizer.<br />

11. Read concentration after it has reached steady state.<br />

12. Correct measured values for the effect of carrier gas used. If AIR no correction required.<br />

If O 2 use following correction factor:<br />

Measured value vol.% =