Repair Information Special Tools Required<br />
3.4 Special Tools Required<br />
PART NUMBER DESCRIPTION SPECIFICATION<br />
1 0070-00-0683 AS3000 Service Manual NA<br />
2 not applicable Vaporizer Instruction Manual NA<br />
3 not applicable Safety Analyzer Dempsey 430 or equivalent<br />
4 not applicable Digital Volt Meter 3 1/2 digit<br />
5 not applicable Agent (and N2O) Analyzer ±0.3 V/V% + 5% of reading<br />
6 0138-00-0012 Test Lung, Adult NA<br />
7 not applicable Digital Pressure Meter BC Biomedical DPM-2301751<br />
NMC Digital Pressure Meter<br />
8 not applicable Central supplied O2, N2O, AIR Minimum of 35 psi, DISS<br />
connections.<br />
9 not applicable Cylinder gases O2, N2O, AIR Full PISS yoke connections.<br />
10 not applicable Hand tools, Allen wrench set Metric<br />
11 0367-00-0080 Tank Wrench NA<br />
12 0103-00-0508 Y-Fitting 15 mm connection<br />
13 0004-00-0076 Respiration Tube (2 required) 0.6 meter silicone, 15 mm<br />
14 0992-00-0139 Breathing Bag 2.3 L silicone<br />
15 0510-00-0020 Krytox Lubricant NA<br />
16 not applicable Vent Tester Certifier-FA PLUS Respiratory<br />
Vent.Tester 4080<br />
17 0997-00-0641 AS3000 Calibration hose NA<br />
18 0103-00-0691 AS3000 test plug NA<br />
19 0040-00-0448 Plug Kit NA<br />
20 0453-00-1216 Regulator Calibration Hose NA<br />
AS3000 Service Manual 0070-10-0683 3 - 3