Warnings and Cautions Repair Information<br />

3.2 Warnings and Cautions<br />

3.2.1 Warnings<br />

3.2.2 Cautions<br />

In the event the instrument covers are removed, observe the following warnings and cautions:<br />

WARNING: Whenever using anesthetic gases, nitrous oxide, oxygen, or<br />

any hospital gas always follow the appropriate agent<br />

evacuation/collection procedures. Use the hospital gas<br />

evacuation system.<br />

CAUTION: This device uses high pressure compressed gas. When<br />

attaching or disconnecting backup gas cylinders, always<br />

turn the cylinder valves slowly. Use the AS3000 flow meters<br />

to bleed down the pressure, watching the cylinder gauge<br />

indicate the depleting cylinder pressure, before<br />

disconnecting the cylinder from the yoke. Always open and<br />

close cylinder valves fully.<br />

CAUTION: This device operates using compressed gas at high<br />

pressures from the hospital central supply. When connecting<br />

gas supply lines attach the hose connection to the machine<br />

before connecting the quick disconnect fitting to the hospital<br />

source. Disconnect the supply hose from the hospital source<br />

connection prior to disconnecting it from the AS3000 gas<br />

connection fittings.<br />

3.3 Troubleshooting Guidelines<br />

1. Identify the problem — Due to the wide variety of potential symptoms, certain<br />

problems may be more subtle than others. Following the guidelines of the tests will help<br />

determine the problem, if one exists.<br />

2. Avoid shorting component leads together — During repair procedures, it can<br />

be tempting to make a quick series of measurements. Always turn the power off before<br />

connecting and disconnecting the test leads and probes. The accidental shorting of<br />

leads can easily stress the components and cause a second failure (aside from the safety<br />

risk).<br />

3. Use the proper equipment — The equipment listed in “Special Tools Required” on<br />

page 3-3, is suggested to fulfill a wide range of troubleshooting requirements. It is<br />

imperative to use the designated equipment in order to ensure proper results of any and<br />

all test procedures.<br />

4. Clean up the repair area — After any repair, clean off the repair area.<br />