28.02.2013 Views

Datascope - Mindray

Datascope - Mindray

Datascope - Mindray

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Periodic Maintenance Functional Tests<br />

6.8.7.6 CMV Child Ventilation Mode Test<br />

1. Attach a breathing circuit and breathing bag.<br />

NOTE: For testing purposes always use a reusable breathing<br />

circuit.<br />

2. Attach an adult test lung to the Y-fitting of the breathing circuit.<br />

NOTE: Limit the volume in the test lung to provide sufficient airway<br />

pressure to satisfy the Low Peak Pressure alarm. Or reduce<br />

the Peak Pressure alarm limit to a lower valve to prevent<br />

the alarm when using an adult test lung.<br />

3. Attach a Vent Tester between the EXP port and the expiratory hose.<br />

4. Set the O 2 flow to 2 L/min and set the N 2 O and AIR flow rates to minimum flow.<br />

5. Set the ventilator controls to:<br />

VENTILATOR CONTROLS VENTILATOR SETTINGS<br />

Patient Type Child<br />

Ventilation Mode CMV<br />

Tidal Volume - VT 120<br />

Breath Rate - freq 20<br />

I:E Ratio - I:E 1:2<br />

Plateau - TP 10<br />

PEEP - PEEP Off<br />

6. Select CMV to begin ventilation.<br />

7. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp. rate<br />

and minute volume values appear on the screen.<br />

8. Verify the Tidal Volume display is within 25ml or ±15% of the delivered volume<br />

(whichever is greater) measured with the Vent Tester within approximately 1 minute from<br />

the start of ventilation.<br />

9. Verify the delivered volume as measured by a Vent Tester at the expiratory port, is within<br />

17% of the Tidal Volume set value within approximately 1 minute from the start of<br />

ventilation.<br />

6.8.7.7 Airway Disconnect Alarm Test<br />

1. While the ventilator is running, disconnect the expiratory limb from the Expiratory Port<br />

on the Breathing System.<br />

2. Verify the following airway pressure disconnect alarm signals activate:<br />

APNEA message appears on the screen.<br />

Low Airway Pressure message appears on the screen.<br />

An alarm tone sounds.<br />

AS3000 Service Manual 0070-10-0683 6 - 29

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!