Periodic Maintenance Functional Tests<br />
FIGURE 6-23 Oxygen Sensor<br />
Calibration Instructions<br />
3. Proceed based on one of the following two conditions:<br />
If the calibration is successful, the screen shown in FIGURE 6-25 will be displayed,<br />
instructing the user to reinstall the oxygen sensor into the Breathing System. Select the<br />
Done button to complete the process.<br />
If the calibration fails, the screen shown in FIGURE 6-26 will be displayed,<br />
instructing the user to either repeat the calibration (by selecting the Repeat Cal button)<br />
or to replace the oxygen sensor. If the oxygen sensor must be replaced, select the Exit<br />
button, replace the oxygen sensor using a push and turn motion, and then repeat the<br />
calibration.<br />
FIGURE 6-25 Oxygen Sensor<br />
Calibration Successful<br />
FIGURE 6-24 Oxygen Sensor Calibration<br />
Progress Bar<br />
FIGURE 6-26 Oxygen Sensor<br />
Calibration Failed<br />
AS3000 Service Manual 0070-10-0683 6 - 23