Periodic Maintenance Functional Tests<br />

FIGURE 6-23 Oxygen Sensor<br />

Calibration Instructions<br />

3. Proceed based on one of the following two conditions:<br />

If the calibration is successful, the screen shown in FIGURE 6-25 will be displayed,<br />

instructing the user to reinstall the oxygen sensor into the Breathing System. Select the<br />

Done button to complete the process.<br />

If the calibration fails, the screen shown in FIGURE 6-26 will be displayed,<br />

instructing the user to either repeat the calibration (by selecting the Repeat Cal button)<br />

or to replace the oxygen sensor. If the oxygen sensor must be replaced, select the Exit<br />

button, replace the oxygen sensor using a push and turn motion, and then repeat the<br />

calibration.<br />

FIGURE 6-25 Oxygen Sensor<br />

Calibration Successful<br />

FIGURE 6-24 Oxygen Sensor Calibration<br />

Progress Bar<br />

FIGURE 6-26 Oxygen Sensor<br />

Calibration Failed<br />

AS3000 Service Manual 0070-10-0683 6 - 23