Installation Guide Assembly<br />

2.2.2.1 Attaching the Mounting Arms and User Interface<br />

(see FIGURE 2-1)<br />

1. Remove the mounting arms from small box and install them onto the AS3000 using a 3<br />

mm allen wrench.<br />

2. Remove the keys from the parts box, and unlock the AS3000 drawers and put parts<br />

box into top drawer.<br />

3. Remove the AGSS and Breathing System mount from the foam packaging, and install<br />

them onto the AS3000.<br />

4. Remove the User Interface and slide it onto the User Interface mounting arm.<br />

5. Apply Loctite 243 to the User Interface cable jackscrews and jackposts. Then, connect<br />

the User Interface cable to the User Interface. Ensure the jackscrews are securely<br />

tightened.<br />

2.2.2.2 Breathing System Connections<br />

(see FIGURE 2-1)<br />

1. Remove the foam and bag from around the Breathing System.<br />

2. Ensure that the serial number tag attached to the bag arm matches the serial number<br />

found on the package carton.<br />

3. Install the Breathing System onto the arm on the left side of the AS3000 by attaching it<br />

to the installed bracket.<br />

4. Attach the Drive Gas Hose to the bottom port of the Breathing System<br />

5. Attach the Heater cable into the port located on the bottom of the Breathing System<br />

6. Install the O 2 Cell to the O 2 Interface<br />

7. Attach the O 2 cell to the O 2 cable and insert the cell into its designated port on the<br />

Breathing System.<br />

8. Attach the Pneumatic Hose Assembly to the designated port on the rear of the Breathing<br />

System.<br />

9. Attach the AGSS Transfer Hose to its designated port on the bottom of the Breathing<br />

System.<br />

2.2.2.3 Anesthesia System Connections<br />

(see FIGURE 2-1)<br />

1. Attach the Fresh Gas Hose of the Breathing System to the CGO port on the system.<br />

AS3000 Service Manual 0070-10-0683 2 - 3