28.02.2013 Views

Datascope - Mindray

Datascope - Mindray

Datascope - Mindray

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Installation Guide Installation Checkout Procedure<br />

16. Proceed based on one of the following two conditions:<br />

If the calibration is successful, the screen shown in FIGURE 2-18 will be<br />

displayed, instructing the user to reinstall the oxygen sensor into the Breathing System.<br />

Select the Done button to complete the process.<br />

If the calibration fails, the screen shown in FIGURE 2-19 will be displayed,<br />

instructing the user to either repeat the calibration (by selecting the Repeat Cal<br />

button) or to replace the oxygen sensor. If the oxygen sensor must be replaced, select<br />

the Exit button, replace the oxygen sensor using a push and turn motion, and then<br />

repeat the calibration.<br />

FIGURE 2-18 Oxygen Sensor<br />

Calibration Successful<br />

17. Verify that the APNEA alarm activates in Manual mode<br />

a. Press the MANUAL/AUTO key.<br />

b. After 60 seconds, verify that:<br />

an alarm tone sounds<br />

the alarm message APNEA is displayed in red text<br />

FIGURE 2-19 Oxygen Sensor<br />

Calibration Failed<br />

18. Verify that CMV ADULT ventilator mode operates and that the tidal<br />

volume display reads within 15% of the CMV set value.<br />

a. Attach a breathing circuit and breathing bag.<br />

NOTE: For testing purposes always use a reusable breathing<br />

circuit.<br />

b. Attach an adult test lung to the Y-fitting of the breathing circuit.<br />

c. Attach a Vent Tester between the EXP port and the expiratory hose.<br />

d. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.<br />

AS3000 Service Manual 0070-10-0683 2 - 19

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!