Setting the ACPGR to - 1, does the following:<br />

� Turns off the display of the ACP graph in trace C.<br />

� Calculates <strong>and</strong> displays the ACP numeric results.<br />

� Removes the GRPH MKR ON OFF softkey from the menu.<br />

� Turns off the ACP graph marker.<br />

ACPGR Adjacent Channel Power Graph On or Off<br />

� If in continuous measurement mode, the numeric results will be updated at the end of every<br />

sweep.<br />

When the graph is generated, the reference level represents an adjacent channel power ratio of<br />

0, <strong>and</strong> the horizontal center represents a frequency offset (<strong>and</strong> channel spacing) of 0 Hz. The<br />

results of ACPGR are not defined for frequency spacings where the graph is drawn below the<br />

bottom graticule line.<br />