TB<br />

Transfer B<br />

Transfers trace B amplitude values from the spectrum analyzer to the controller.<br />

Syntax<br />

Related Comm<strong>and</strong>s: MDS, TA, TDF.<br />

Example<br />

DIM A(4011<br />

OUTPUT 718;"IP;"<br />

OUTPUT 718;"SNGLS;CF 300MHZ;SP 2MHZ;TS;"<br />

OUTPUT 718;"TDF P;TB;"<br />

FOR N = 1 TO 401<br />

ENTER 718;A(N)<br />

NEXT N<br />

FOR N = 1 TO 401<br />

PRINT A(N)<br />

NEXT N<br />

Description<br />

The operation of TB is similar to the operation of TA.<br />

TB Transfer B<br />

&serves memory area for array.<br />

Initializes analyz4x<br />

Activates single-sweep mode, changes center<br />

frequency <strong>and</strong> span, takes sweep.<br />

Changes trace data format, outputs trace<br />

B<br />

FOR NEXT loop m.oves each element of<br />

trace A to the computer:<br />

FOR NEXT loop moves the trace values<br />

from the computer to the printer:<br />

Prints out the results.<br />

See Chapter 3, “Different Formats for Trace Data Transfers, n for more information about<br />

transferring trace data.<br />

Programming Comm<strong>and</strong>s 5-543