09.12.2012 Views

HP 8590 E & L Series Spectrum Analyzers and HP 8591C Cable TV ...

HP 8590 E & L Series Spectrum Analyzers and HP 8591C Cable TV ...

HP 8590 E & L Series Spectrum Analyzers and HP 8591C Cable TV ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

DL<br />

Display Line<br />

DL Display Line<br />

Defines the level of the display line in the active amplitude units <strong>and</strong> displays the display line<br />

on the spectrum analyzer screen.<br />

Syntax<br />

Item Description/Default Range<br />

Number Any real or integer number. Default units are dBm. Dependent on the<br />

reference level.<br />

Equivalent Softkey: l3SP LINE OM OFF .<br />

Preset State: DL OFF<br />

Step Increment: 1 major division.<br />

Related Comm<strong>and</strong>s: AUNITS, AUTO, LG, LN, NRL, RL, ROFFSET, TH.<br />

Example<br />

OUTPUT 718;"AUNITS DBM;" Changes the active amplitude units to dBm.<br />

OUTPUT 718;"DL ON;" Turns on the displa?/ line.<br />

OUTPUT 718;"DL -5DM;" Changes display line to -5 dBm.<br />

Description<br />

Activating video trigger mode activates the display line. The AUTO comm<strong>and</strong> <strong>and</strong> “DL OFF; ”<br />

turn off the display line. See “AUNITS” for more information on changing the active amplitude<br />