AMBPL<br />

Trace A Minus Trace B Plus Display Line<br />

Subtracts trace B from trace A, adds the display line value to the difference, <strong>and</strong> sends the<br />

result to trace A during every sweep of the spectrum analyzer.<br />

Syntax<br />

Equivalent Softkey: NORMLIZE ON OFF .<br />

Preset State: AMBPL OFF.<br />

Related Comm<strong>and</strong>s: ADD, AMB, CONTS, CLRW, DL, MXMH, SNGLS, SUB, TS, VAVG, VIEW.<br />

Restrictions: Not compatible with Analog+ display mode. See “ANLGPLUS” for more<br />

information.<br />

Example<br />

10 OUTPUT 718;"IP;SNGLS;RL 20DB;"<br />

20 OUTPUT 718;"MOV TRA,5000;"<br />

30 OUTPUT 718;"VIEW TRA;"<br />

40 OUTPUT 718;"MOV TRB,4000;"<br />

50 OUTPUT 718;"VIEW TRB;"<br />

60 OUTPUT 718;"DL ODM;”<br />

70 OUTPUT 718;"AMBPL ON;"<br />

80 OUTPUT 718;"BLANK TRB;VIEW TRA;"<br />

90 END<br />

5-60 Programming Comm<strong>and</strong>s<br />

OFF<br />

ON<br />

/<br />

Initializes spectrum analym activates singlesweep<br />

mode.<br />

Sets trace A to 5000 meoxurement units, which<br />

is equal to -10 dBm.<br />

Sets trace B to 4000 measurement units, which<br />

is equal to -20 dBm.<br />

Sets display line to 0 dBm, which is at 6000<br />

measurement units.<br />

&forms trace A - trace B + display line. The<br />

result is 5000 - 4000 + 6000 = 7000 or 10 dBm.<br />

Note that this has resulted in a subtraction of<br />

amplitude in dBm, -10 dBm -(-20 dBm) = 10<br />

dBm.