Description<br />

RCEI’ Recall Trace<br />

The state <strong>and</strong> trace data are recalled when the trace destination is trace A, trace B, or trace C.<br />

When using a user-defined trace or a trace range for the trace destination, only the trace data<br />

is recalled.<br />

When recalling frequency-amplitude correction data, you need to specify AMPCOR as the<br />

destination. When recalling limit line table data, specify LIMILINE as the destination.<br />

To avoid overwriting the recalled trace data, the VIEW comm<strong>and</strong> should be performed<br />

immediately after the RCLT comm<strong>and</strong> when recalling trace data (see line 40 in the example).<br />

Note The RCLT recalls trace data from spectrum analyzer memory. See “LOAD” or<br />

“SAVRCLN” to recall trace data from the memory card.<br />

Programming Comm<strong>and</strong>s 5-461