09.12.2012 Views

HP 8590 E & L Series Spectrum Analyzers and HP 8591C Cable TV ...

HP 8590 E & L Series Spectrum Analyzers and HP 8591C Cable TV ...

HP 8590 E & L Series Spectrum Analyzers and HP 8591C Cable TV ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

FFTCONTS<br />

FFT Continuous Sweep<br />

Performs a fast Fourier transform (FFT) continuously on the current signal.<br />

Syntax<br />

Equivalent Softkey: CONTINUS FFT .<br />

Prerequisite Comm<strong>and</strong>s: LN, Sl?<br />

Related Comm<strong>and</strong>s: FFTOFF, FFTSNGLS.<br />

Example<br />

OUTPUT 718; “SP OHZ; ” Sets the spectrum analyzer to zero span.<br />

OUTPUT 718; “LN; ” Changes the amplitude scale to linear:<br />

OUTPUT 718;"FFTCONTS;" Starts the continuous FlT<br />

Description<br />

FFT.CONTS FFT Continuous Sweep<br />

You should change the frequency span to 0 Hz <strong>and</strong> the amplitude scale to linear before<br />

executing FFTCONTS. When FFTCONTS is executed, the spectrum analyzer does the following:<br />

� Changes to the continuous sweep mode.<br />

� If the current detector is the peak detector, changes to the sample detector.<br />

Does an FFI’ on trace A. Trace A is then placed in both the clear-write <strong>and</strong> store-blank<br />

modes. (When the spectrum analyzer is in both the clear-write <strong>and</strong> store-blank modes, the<br />

trace data is still taken from the spectrum analyzer input during every measurement sweep,<br />

but the trace is not shown on the spectrum analyzer display.)<br />

� Places the results in trace B (trace B is placed in the view mode).<br />

You can use the results of the FFTCLIP comm<strong>and</strong> to determine if the FFI’ data is valid. If you<br />

want to view the input data (trace A), execute “TRDSP TRA,ON; ‘I.<br />

Restrictions<br />

Executing FFTCONTS turns off the following functions: windows display mode (WINON),<br />

N dB point measurement (NDBPNT), gate utility functions (GDRVUTIL), TO1 measurement<br />

(TOI), marker table (MKTBL), peak table (PKTBL), percent AM (PCTAM), peak zoom<br />

(PKZOOM), power menu measurements (ACP, ACPE, C<strong>HP</strong>, <strong>and</strong> OBW), or Analog+ display mode<br />

(ANLGPLUS).<br />

FFTCONTS uses the following when performing an FFT:<br />

� The flat top filter window. See “FFT” for more information about the flat top filter window.<br />

w Trace B <strong>and</strong> trace C. If you want to save the trace data that is in trace B or trace C, you<br />

should save the trace data before executing FFTAUTO, FFTCONTS, or FFTSNGLS. (See<br />

“SAVET” for more information about saving trace data.)<br />

� The FFTMKR comm<strong>and</strong> to turn on the FFI markers (you can use the MKA or MKF to<br />

determine the amplitude or frequency of the marker).<br />

You should execute the FFTOFF comm<strong>and</strong> when you are finished using the FFT measurement.<br />

Programming Comm<strong>and</strong>s 5-167