ACPPAR ACP Manual or Auto<br />

‘able 5-5. <strong>Spectrum</strong> Analyzer Settings, ACPPAR is Set to Automatic<br />

Measurement Res Video Span Step Sweep<br />

B<strong>and</strong>width B<strong>and</strong>width Size Time<br />

Adjacent Channel Power Highest setting that does 10 x RB ACPSP Auto<br />

WJ’)<br />

not exceed 0.025 x<br />

(400 x ACPSP)<br />

(4OOxACPSP)<br />

ACPBW’ INT[< ~xACPSP+I.~XACPBW)~<br />

Channel Power (C<strong>HP</strong>) Highest setting that does 10 x RB 2 x ACPBW ACPSP Auto<br />

not exceed 0.025 x<br />

ACPBW’<br />

Occupied B<strong>and</strong>width Highest setting that does 10 x RB 3 x ACPSP ACPSP Auto<br />

VW<br />

not exceed 0.02 x<br />

ACPSP *<br />

* If Option 130 is not installed in the spectrum analyzer, the narrowest resolution b<strong>and</strong>width is limited to 1 kHz.<br />

When ACPPAR is set to 0, you must set the spectrum analyzer settings for the ACP, ACPE, C<strong>HP</strong>,<br />

or OBW measurement. You must ensure that trace A contains the RF spectrum to be measured,<br />

<strong>and</strong> that the frequency span, resolution b<strong>and</strong>width, video b<strong>and</strong>width, <strong>and</strong> the detector are set<br />

appropriately for the measurement. When ACPPAR is set to 0, the measurement comm<strong>and</strong>s<br />

(ACP, ACPE, OBW, C<strong>HP</strong>) do not take a sweep before making the measurement.<br />

You can execute the ACPPAR comm<strong>and</strong> two different ways. You can either execute the<br />

ACPPAR comm<strong>and</strong> directly (for example, “ACPPAR 1; ‘I) or use the MOV comm<strong>and</strong> to move<br />

the 1 or 0 into the ACPPAR comm<strong>and</strong> (for example, “MOV ACPPAR, 1; ‘I). If you use the MOV<br />

comm<strong>and</strong>, no text is displayed in the active function area during comm<strong>and</strong> execution.<br />

Query Response<br />

r manual 7<br />

A<br />

0<br />

/ �<br />

output<br />

termination ---)<br />

Programming Comm<strong>and</strong>s 5-45