ACPMK<br />

Adjacent Channel Power Marker On or Off<br />

ACPMK Adjacent Channel Power Marker On or Off<br />

Determines if the graph marker function is enabled or disabled for the adjacent channel power<br />

(ACP) graph.<br />

Syntax<br />

Equivalent Softkey: GRPH ON OFF .<br />

Related Comm<strong>and</strong>s: ACPGR, ACPGRAPH, C<strong>HP</strong>GR, ACPSNGLM, ACPCONTM.<br />

Example<br />

OUTPUT 718;"ACP<br />

OUTPUT 718;"MOV<br />

OUTPUT 718;"MOV<br />

Description<br />

. II<br />

8<br />

&forms the adjacent channel power measurement.<br />

xacpmk<br />

ACPGR,l;" Enables the ACP graph function <strong>and</strong> creates a graph of<br />

ACP as a function of the frequency offset from the center<br />

freq-y.<br />

ACPMK,l;" Enables the ACP graph mm-km<br />

Setting ACPMK to 1, does the following for ACP graph:<br />

w Positions a reference marker on the trace for ACP graph at the center frequency position.<br />

w Positions a A marker on the trace at one channel spacing above the center frequency.<br />

� Displays numeric values for delta frequency, ACP ratio, <strong>and</strong> channel power at the marker<br />

position.<br />

Setting ACPMK to 1, does the following for channel power graph:<br />

w Positions a marker on the trace for channel power graph at center frequency position.<br />

� Displays numeric values for frequency <strong>and</strong> channel power at the marker position.<br />

Setting the ACPMK to -1, turns off the marker function.<br />

Programming Comm<strong>and</strong>s 5.43