ONSRQ<br />

On Service Request<br />

Executes the list of analyzer comm<strong>and</strong>s whenever a service request occurs.<br />

Syntax<br />

Delimiter<br />

Item<br />

Analyzer comm<strong>and</strong><br />

comm<strong>and</strong>s to<br />

- be executed 7<br />

delimiter delimiter<br />

Description/Default Rmife<br />

XONSRO<br />

Matching characters marking the beginning <strong>and</strong> end of the list of - 1 \ @ = / ^ $ % ; ! ’ :<br />

spectrum analyzer comm<strong>and</strong>s. ” &<br />

Any spectrum analyzer comm<strong>and</strong>.<br />

Related Comm<strong>and</strong>s: CAT, DISPOSE, ERASE, IP, ONDELAY, ONEOS, ONMKR, ONSWP,<br />

ONTIME, SRQ.<br />

Example<br />

OUTPUT 718;"ONSRQ !PU;PA 100,lOO;TEXT QSRQ OCCURRED@; !;” “SRQOCCURRED”isdisplayed<br />

on the spectrum<br />

analyzer screen if an SRQ<br />

is encountered.<br />

Description<br />

Limit the number of characters (between the delimiters) in the list of spectrum analyzer<br />

comm<strong>and</strong>s to a maximum of 2047 characters.<br />

Clearing the ONSRQ definition: IP clears the ONSRQ definition. You can use the DISPOSE<br />

comm<strong>and</strong> to clear the ONSRQ definition also.<br />

5-402 Programming Comm<strong>and</strong>s