IB<br />

Input B<br />

Provides a method for putting values into trace B. The spectrum analyzer expects 401 two-byte<br />

integers. The data values can represent the range of integer numbers.<br />

Syntax<br />

Item Description/Default Range<br />

Data byte &bit byte containing numeric or character data.<br />

Example<br />

10 ASSIGN @a TO 718;FORMAT ON<br />

20 ASSIGN BSa-bin TO 718;FORMAT OFF<br />

30 INTEGER Binary(l:401)<br />

40 OUTPUT (PSa;"CF 300MZ;CLRW TRB;BLANK TRA;SP<br />

1OMZ;SNGLS;TS;"<br />

50 OUTPUT QISa;"TDF B;MDS W;TB;"<br />

60 ENTER QSa-bin;Binary(*)<br />

70 OUTPUT @Sa;"CF 1OOMZ;RB 30KZ;SP 1MZ;TS;"<br />

80 DISP "PRESS CONTINUE WHEN READY"<br />

90 PAUSE<br />

100 OUTPUT OSa;"IB";<br />

110 OUTPUT aSa_bin;Binary(*)<br />

120 OUTPUT 718;"VIEW TRB;"<br />

130 END<br />

Description<br />

Llimensions an array called “‘Binary. ”<br />

Zzkes a measurement sweep.<br />

Outputs trace B (in binary) to computer:<br />

Stores trace data in array.<br />

Changes the spectrum analyzer settings.<br />

Prepares spectrum analyzer to receive<br />

trace B data stored in array.<br />

Sends trace B data to spectrum analyzer<br />

The IB comm<strong>and</strong> sends trace B data as binary data only; IB is independent of the measurement<br />

data size (MDS) comm<strong>and</strong>. The IB comm<strong>and</strong> expects 802 data bytes (401 data points, two bytes<br />

each).<br />

5.234 Programming Comm<strong>and</strong>s