MKSP<br />

Marker to Span<br />

Sets the start <strong>and</strong> stop frequencies to the values of the delta markers.<br />

Syntax<br />

Equivalent Softkey: MKl3 B -> SPAN.<br />

Related Comm<strong>and</strong>s: MKD, SI?<br />

Example<br />

10 OUTPUT 718;"IP;"<br />

20 OUTPUT 718*"MKMIN-"<br />

30 OUTPUT 718;"MKD;"'<br />

40 OUTPUT 718;"MKPK HI;"<br />

50 OUTPUT 718;"MKSP;"<br />

60 END<br />

Description<br />

MKSP<br />

MKSP Marker to Span<br />

Initializes spectrum analyzer<br />

Places a marker at the minimum amplitude of trace.<br />

Activates marker delta.<br />

Places marker at highest amplitude of trace.<br />

Changes span to the values of the left <strong>and</strong> right markers.<br />

The left marker specifies start frequency, <strong>and</strong> the right marker specifies stop frequency. If MKD<br />

is off, no operation is performed.<br />

Programming Comm<strong>and</strong>s 5-357