GDRVGDEL<br />

Gate Delay for the Frequency Window<br />

GDRVGDEL Gate Delay for the Frequency Window<br />

For the frequency window only, GDRVGDEL sets the time delay from when the gate trigger<br />

occurs to when the gate is opened.<br />

Syntax<br />

\ / XGDRVGDEL<br />

Item I Description/Default R=u3e<br />

Number Any real or integer number. Default unit is seconds. 1 ps to 65 ms.<br />

Equivalent Softkey: GDRVGDEL is equivalent to GATE DELAY when using the gate utility<br />

(GATE UTILfTY ) functions.<br />

Option Required: Option 105. Option 101 is recommended.<br />

Preset Value: 1 ps<br />

Related Comm<strong>and</strong>s: GDRVUTIL, GDRVGLEN, GD, GL.<br />

Example<br />

OUTPUT 718;"MOV GDRVGDEL,lUS;" Sets the gate marker delay to 1 ps.<br />

Description<br />

If the frequency window is currently the active window, GDRVGDEL updates the gate position<br />

markers <strong>and</strong> the position of the gate. If the time window is currently active, only the gate<br />

position markers are updated.<br />

You can execute the GDRVGDEL comm<strong>and</strong> two different ways. You can either execute the<br />

GDRVGDEL comm<strong>and</strong> directly (for example, "GDRVGDEL IMS ; ‘I) or use the MOV comm<strong>and</strong><br />

to move the value for the time delay into the GDRVGDEL comm<strong>and</strong> (for example, “MOV<br />

GDRVGDEL, 1MS ; ‘I). If you use the MOV comm<strong>and</strong>, no text is displayed in the active function<br />

area during comm<strong>and</strong> execution.<br />

The GDRVGDEL comm<strong>and</strong> is a gate utility function (see “GDRVUTIL” for more information<br />

about the gate utility functions). Executing a gate utility comm<strong>and</strong> without Option 105<br />

installed in the spectrum analyzer generates the COMMAND ERROR: message <strong>and</strong> an SRQ 140.<br />