LIMIFT<br />

Select Frequency or Time Limit Line<br />

Selects how the limit-line segments are defined: according to frequency, or according to the<br />

sweep time setting of the spectrum analyzer.<br />

Syntax<br />

Equivalent Softkey: LIMIFT is equivalent to LIMITS FRIJ TIME.<br />

Related Comm<strong>and</strong>s: LIMIDEL, LIMILINE, LIMIMODE, LIMIREL, LIMISEG, LIMISEGT,<br />

SEGDEL, SENTER, SENTERT.<br />

Example<br />

OUTPUT 718;"LIMIFT TIME;" Ifthe current limit-line table is a frequency limit-line table,<br />

it is purged. LIMIFT TIME places the limit-line segments<br />

on the spectrum analyzer display with respect to the sweep<br />

time of the spectrum analyzer:<br />

Description<br />

If you execute “LIMIFT TIME;“, LIMISEGT, or SENTERT, the limit-line segments are placed on<br />

the spectrum analyzer display with respect to the sweep time setting of the spectrum analyzer.<br />

If you execute “LIMIFT FREQ;“, LIMISEG, or SENTER, the limit-line segments are placed<br />

according to the frequency that is specified for each segment. If a limit line has already been<br />

defined, changing the LIMIFI’ setting clears the existing limit line.<br />