SENTERT Segment Entry for Sweep Time Limit Lines<br />

Item Description/Default Range<br />

Uumber Any real or integer number. For amplitude, the The range for the<br />

default unit is dBm. For sweep time, the default amplitude varies<br />

unit is seconds. with ROFFSET.<br />

The range for the<br />

sweep time is the<br />

sweep time range<br />

of the spectrum<br />

analyzer.<br />

User-defined variable A variable defined by VARDEF or ACTDEF Any valid<br />

comm<strong>and</strong>s. variable name.<br />

?redefined variable A comm<strong>and</strong> that acts as a variable. Refer to ‘Ikble<br />

5-1.<br />

Trace element An element of trace A, trace B, trace C, or a<br />

user-defined trace.<br />

Related Comm<strong>and</strong>s: LIMIMODE, LIMIREL, LIMISEGT, SEGDEL, SENTER.<br />

Example 1<br />

This example enters limit-line values into the upper <strong>and</strong> lower limit-line tables.<br />

OUTPUT 718;"RL -1ODB;"<br />

OUTPUT 718~"LIMIDEL*"<br />

OUTPUT ~~~I~~LIMIFT +IME;"<br />

Sets the reference level to -10 dB<br />

Erases any the current limit line table.<br />

Sets the limit lines to be based on<br />

sweep time.<br />

OUTPUT 718;"LIMIMODE UPLOW;LIMIREL ON;" Specifies the upper <strong>and</strong> lower limitline<br />

table as rektive.<br />

OUTPUT 718;"SENTERT lOMS,-lODB,-50DB,FLAT;" Enters in values for a segment.<br />

OUTPUT 718;"SENTERT OMS,-15DB,-GODB,FLAT;" Enters in values for a segment.<br />

OUTPUT 718;"LIMITEST 0N;TS;" Turns on the limit-line testing.<br />

Example 2<br />

OUTPUT 718;"LIMIFT TIME;" Sets the limit lines to be based on sweep<br />

time.<br />

OUTPUT 718;"LIMIMODE DELTA;LIMIREL OFF;" Spec@..es the mid <strong>and</strong> delta table format<br />

<strong>and</strong> jkzd type.<br />

OUTPUT 718;"SENTERT lOMS,-20DB,lODB,FLAT;" Enters in values for a segment.<br />

OUTPUT 718;"SENTERT OMS,-30DB,20DB,FLAT;" Enters in values for a segment.<br />

OUTPUT 718;"LIMITEST 0N;TS" Turns on the limit-line testing.<br />

Description<br />

Each limit-line segment is specified with a starting sweep time, an upper or median amplitude<br />

value, a lower or delta amplitude value, <strong>and</strong> a segment type. The segment type defines how the<br />

line segment is to extend from its starting point to the next segment.<br />

Programming Comm<strong>and</strong>s 5488