Test Program<br />

To test the connection between the computer <strong>and</strong> the spectrum analyzer, turn on your<br />

spectrum analyzer <strong>and</strong> follow the instructions below.<br />

1. Your <strong>HP</strong> 9000 <strong>Series</strong> 300 computer may have either a soft-loaded or built-in language<br />

system. If your language system is built-in, remove any disks from the drives <strong>and</strong> turn on<br />

the computer.<br />

2. If your language is soft-loaded, install the BASIC language disk into the proper drive. Turn<br />

the computer power on. After a few seconds, the BASIC READY message appears; the<br />

computer is now ready for use.<br />

For further information on loading BASIC on your system, consult your BASIC manual.<br />

Check the <strong>HP</strong>-IB address of the spectrum analyzer: press @K@, More 1 of 3 ,<br />

ANALYZER ADDRESS . The usual address for the spectrum analyzer is 18. If necessary, reset<br />

the address of the spectrum analyzer: press (ml, More 1 of 3 , ANALYZER ADDRESS ,<br />

1Hz) (or enter the appropriate address).<br />

4. Enter the following program, then press IRUN) on the computer. If you need help entering<br />

<strong>and</strong> running the program, refer to your computer <strong>and</strong> software documentation.<br />

The program shows that the computer is able to send instructions to, <strong>and</strong> read information<br />

from, the spectrum analyzer.<br />

10 PRINTER IS 1<br />

20 Analyzer=718<br />

30 CLEAR Analyzer<br />

40 OUTPUT Analyzer ; ; SNGLS ; ”<br />

50 OUTPUT Analyzer ; “CF 300MZ ; TS ; ”<br />

60 OUTPUT Analyzer ; “CF? ; I’<br />

70 ENTER Analyzer;A<br />

80 PRINT “CENTER FREQUENCY = ;A; “Hz” ;<br />

90 END<br />

The program tells the spectrum analyzer to perform an instrument preset <strong>and</strong> enter<br />

single-sweep mode. Next, the program sets the center frequency to 300 MHz <strong>and</strong> takes a<br />

sweep.<br />

CENTERFREQUENCY= 3.OE+8Hz.<br />

If the computer does not display the center frequency, refer to “If There is a Problem” at the<br />

end of this chapter.<br />

1-6 Preparing for Use